The information to be found in this section is intended for the exclusive use of healthcare professionals.
A fully automated system
|
|
|
|
A.
|
A thermo-regulated carousel set at 37°C with 90 positions for disposable cuvettes. A spectrophotometer for measuring absorbance in certain immunoassay reactions is incorporated into the carousel
|
B.
|
A pipetting arm that pipettes both reagents and samples
|
C.
|
A sample tray with 64 positions for samples, calibrators and controls
|
D.
|
A refrigerated reagent compartment, consisting of 15 rails containing the immunoassay reagent racks. The reagent cartridges are stored in the refrigerated reagent compartment between 12 and 15°C while the analyzer is operating, and between 8 and 10°C in standby mode
|
E.
|
An automatic cuvette loader, holding 960cuvettes at a time (pre-formed as a cube)
|
F.
|
Two barcode readers (one located on the front face of the analyzer, the other integrated into the reagent compartment) identify samples, ancillary reagents and reagent cartridges
|
G.
|
Four washers for washing magnetic particles
|
H.
|
A luminometer measuring luminescence in immunoassay reactions (directly in the cuvette)
|
|
|
AVAILABLE TESTS