The information to be found in this section is intended for the exclusive use of healthcare professionals.
Connective tissue diseases (CTD): Zenit RA Centromere B
Code: 41431
Parameter/assay principle: Quantitative detection of IgG antibodies to centromere B
Number of tests: 50
Sample type: Serum or plasma (EDTA, Heparin)
Sample volume: 4 µl
Incubation time: 10 + 10 minutes
Stability of calibration: 3 weeks
Cartridge stability on board: 8 weeks
Measuring range: 0-640 AU/mL
Results (AU/mL): Negative <10 | Positive ≥10
Clinical specificity: -
Clinical sensitivity: -
Limit of detection: 7.2 AU/mL
Precision: Intra-assay CV% ≤9.8 | Inter-assay CV% ≤8.8
Relative specificity: § 98.1%
Relative sensitivity: § 97.9%
*The content of this file is for informational purposes only. The latest, updated version of the IFU, to which you should always refer, is the one accompanying the product and is to be found on the disk inside the product kit